High resolution oscilloscopes and FFT spectrum analyzers.

ADC-212 and ADC-216 high resolution oscilloscopes Over the last few years there have been some major changes in the oscilloscope market. Analog scopes have given way to digital storage oscilloscopes (DSO's) and more recently PC based oscilloscopes have become an alternative to conventional 'benchtop' oscilloscopes.

The driving force behind DSO's has been digital electronics. As clock speeds in digital circuits increase, manufacturers have responded by designing scopes with faster sampling rates and higher bandwidths.

Unfortunately analogue designers have been left behind: the quest for higher speed has been at the expense of accuracy, dynamic range and precision.

What is a 'Precision' DSO?

In digital electronics a signal change of 1% is usually no problem, but in audio electronics 0.1% distortion or noise can be a disaster. Conventional DSO's only offer 8 bit resolution (8 bit ADC), so can detect at best a 0.4% signal change (see table). This means that they are unsuitable for many applications ranging from monitoring sensors and transducers (temperature, current, pressure) to detecting noise and vibration. Precision oscilloscopes require more than 8 bit resolution.

Oscilloscope resolution No of steps Smallest change that can be detected Dynamic range
6 bit oscilloscopes 64 1.6% (16000ppm) 36dB
8 bit oscilloscopes 256 0.39% (4000ppm) 54dB
12 bit oscilloscopes 4096 0.024% (244ppm) 72dB
16 bit oscilloscopes 65536 0.0015% (15ppm) 96dB

As well as resolution noise is an issue. The amplifiers that make up the 'front end' of a conventional DSO are designed to have a high bandwidth. Low noise is not a priority and so 1% noise figures are typical. The designer of a 16 bit oscilloscope has a tough job, noise must be kept below 0.0015% (15ppm), otherwise the extra resolution will be swallowed up by noise.

Accuracy is not usually regarded as important for most oscilloscopes. You can make measurements within a few percent (most DSO's quote 3% to 5% DC accuracy) but for accurate measurements you have to reach for a multimeter. With precision oscilloscopes, accurate measurements are possible at full speed. Most hand held meters are 12 bit resolution and a 16 bit oscilloscope is equivalent in resolution to a 4.5 digit 'benchtop' meter.

Precision oscilloscopes in use

The oscilloscope displays below all show the output of a high quality audio pre amp being fed with a pure sinewave. The output is fairly clean, but has a picked up a small amount of 50Hz mains noise. The first display is an 8 bit oscilloscope (almost all DSO's are 8 bit), the second shows a 12 bit oscilloscope (Pico ADC-212) and the third a 16 bit oscilloscope (Pico ADC-216).

8 bit oscilloscope example

An 8 bit oscilloscope gives a good visual representation of the wave. From the top left hand view you can clearly see that it is a sine wave and work out its approximate frequency and amplitude. If however, you wish to 'zoom' in on the signal to magnify an area of interest, this soon shows up the limitations of an 8 bit oscilloscope. The bottom left trace shows the top trace magnified by x50.

The view on the right shows a spectrum analyser plot (FFT) of the signal. The peak shows the frequency to be exactly 200Hz. The other signals (approx 60dB down) represent the 'noise floor' of the 8 bit scope. To put this into context a typical tape recorder has about 50 to 60dB of dynamic range. The signal contains 50Hz noise due to mains pickup, but the 8 bit scope does not have enough resolution to detect it.

8 bit resolution oscilloscopes

12 bit oscilloscope example

The same signal captured with a 12 bit oscilloscope (Pico ADC-212) looks the same in the normal scope view. The x50 view looks much better, the digitisation 'steps' are only just visible. On the spectrum analyser, we still cannot see the 50Hz noise, but we know that there is at least a 80dB difference between the signal and the noise. For most audio applications, this is enough (near CD quality) to tell that any noise will not be audible.

12 bit resolution oscilloscopes

16 bit oscilloscope example

With 16 bit oscilloscopes (Pico ADC-216), the x50 trace is smooth, with no signs of distortion caused by the digitisation. The spectrum trace shows 50Hz noise (approx 86dB down) and also harmonics of 50Hz (at 150Hz and 250Hz). To test that the 50Hz noise was coming from the amplifier not the sinewave source we connected the sinewave directly to the ADC-216: the 50Hz peak disappeared.

16 bit resolution oscilloscopes


DrDAQ data logger Pico Technology Company Information Ordering Information Technical support Download oscilloscope and data acquisition software Latest news about oscilloscopes and data loggers Data Acquisition Oscilloscopes and spectrum analysers Pico Technology Home Page oscilloscopes and data loggers